Procedimentos de Verificação e Calibração em Ensaios Não Destrutivos
The checking of equipment should be done at intervals specified by standard and recorded. The records should be kept in an appropriate place and available for consultation. Calibration of equipment should be in accordance with standards used and traceable to a primary calibration system.
Black Light
Black light filters and reflectors shall be checked daily for cleanliness and integrity. Dirty or damaged reflectors and filters shall be replaced or repaired as appropriate.
he black light intensity shall be at an acceptable minimum intensity of 1000 μW/cm 2 on the surface to be examined Any intervention in the equipment such as lamp replacement, electrical repair, etc; shall be checked for light intensity. The periodicity of the verification should be in accordance with the standard used for inspection.
Examples: ASME SEC V article 24 SE 165 - weekly; ASTM E 1417 - daily; ASME SEC V - article 6 T676.4d - every 8 hours
Light Meters
White light and black light meters shall be calibrated in accordance with the standard used for testing. Examples: ASME SEC V article 6 T660 - annually; ASTM E 1417 - semi-annually.
Pressure / Operating temperature of water for washing
Indicators and controls should be reviewed at the beginning of each shift to ensure proper adjustment. Indicators and controls should be calibrated in accordance with the requirements of the specific standard.
Calibration of Drying Oven
Temperature shall be controlled with a calibrated device capable of maintaining oven temperature at 71°C maximum. Calibration should be performed quarterly.
Cleaning of the Testing Area
The test area should be clean and free from excessive penetrant contamination. The check should be performed daily.